Original Article

A Novel Automated Endovascular Variable Aortic Control Device to Expand Function of Standard REBOA Catheters

  • Timothy Williams Wake Forest School of Medicine

Abstract

Abstract


Background


Endovascular methods for hemorrhage control, including Resuscitative Endovascular Balloon Occlusion of the Aorta (REBOA), are evolving and are increasingly being applied clinically. Partial flow strategies to mitigate the consequences of complete aortic occlusion have been demonstrated in pre-clinical models to enhance REBOA and expand its application to various shock states. Initial studies demonstrated that controlled partial flow requires precision beyond the capabilities of manual balloon volume adjustment, therefore automation is required. Our group previously developed a proof-of-concept computer-controlled extracorporeal flow circuit capable of precision aortic flow regulation, but it was not clinically applicable. To bring this concept closer to clinical applicability, we have developed the first endovascular strategy to achieve precision aortic flow regulation, termed Endovascular Variable Aortic Control (EVAC).


Methods


Following instrumentation, 5 Yorkshire-cross swine were subjected to controlled 25% hemorrhage, followed by precision low volume aortic flow regulation using a commercially available compliant balloon catheter pre-positioned in the descending thoracic aorta, connected to a custom, wireless syringe pump. Closed-loop feedback algorithms based on streaming physiologic data were used to determine balloon volume changes.


Results


The EVAC syringe pump was highly effective at maintaining precise aortic flow throughout the 45-minute intervention period during steady state conditions as well as during rapid fluid administration. Aortic flow and distal mean arterial pressure remained stable during EVAC, despite changing proximal hemodynamics. Balloon volume was dynamic, averaging over 500 changes during intervention, with a mean volume change of 6.7 uL and a maximal change of 100 uL.


Conclusion


The EVAC syringe pump is capable of achieving aortic flow regulation with high precision, beyond what is achievable with manual control. This serves as a model for future device design, enabling as-of-yet unachievable clinical therapies for hemorrhage and shock states. Future technological development is required to fully translate this into clinical use.


Level of Evidence – Level V


Study Type – Translational Science


Keywords – EVAC, P-REBOA, REBOA, automation, hemorrhage


 


 


 


 

Published

2019-01-02

How to Cite

WILLIAMS, Timothy. A Novel Automated Endovascular Variable Aortic Control Device to Expand Function of Standard REBOA Catheters. Journal of Endovascular Resuscitation and Trauma Management, [S.l.], v. 3, n. 1, jan. 2019. ISSN 2002-7567. Available at: <http://www.jevtm.com/journal/index.php/jevtm/article/view/65>. Date accessed: 27 may 2019. doi: https://doi.org/10.26676/jevtm.v3i1.65.